Overview
The Canon CXDI-50C portable, lightweight Digital Radiography system provides high quality, filmless image capture for a broad range of radiographic applications, including trauma, ICU, and bedside exams. Featuring proprietary, advanced Canon LANMIT (Large Area New-MIS Sensor and TFT) detector technology, the Canon CXDI-50C Digital Radiography system delivers high-quality diagnostic images efficiently with minimal x-ray exposure to patients, making it ideal for pediatric and orthopedic use. The versatile CXDI-50C can upgrade your existing radiographic suite or can be retrofit to an upright stand, universal stand, or imaging table. The CXDI-50C can also be retrofit to certain portable radiographic units.
Features
Large Imaging Area in a Compact, Portable Design
The large 14 inch x 17 inch imaging area is provided by a detector that is just under an inch thick (0.9 inches) and weighs only 10.6 pounds. The CXDI-50C is especially useful with patients who have limited mobility and for capturing images at angles that are difficult to acquire with a detector in an installed table or wall stand.
Immediate Results
An on-screen preview image is available just 3-5 seconds after exposure. A monitor (not included) can be added for viewing high-resolution images in the emergency department or radiography suite.
LANMIX MLT(S) Image Processing
In addition to Canon’s advanced noise reduction processing, MLT(S) image processing utilizes the multiple frequencies within the image data to emphasize edges and display appropriate dynamic range. Images of extremities, for example, show the subtle details of trabecular bone structure and soft tissue in high diagnostic accuracy. This process can be preset for repeatability as well as adjusted post-acquisition for variances in diagnostic requirements.
Click here for a larger image
|
|
Enhanced Productivity and Program Imaging Mode
Program Imaging Mode for Canon CXDI Digital Radiography control software with DICOM Modality Worklist assists in optimizing workflow in the imaging suite. This high performance solution utilizes the Procedure Code ID from the RIS to configure the exam on the CXDI Digital Radiography system.
Canon’s LANMIT® Technology
Amorphous Silicon (a-Si) Flat Panel Sensor (LANMIT®) consists of a precision amorphous silicon metal-insulator semiconductor (MIS) sensor and Thin Film Transistor (TFT) array. LANMIT® flat panel sensors produce a highly stable output signal. The result is a family of reliable and durable sensors that provide the highest quality images without the time and costs involved with film or cassette handling. A 14-bit analog-to-digital conversion process converts the raw image to digital form producing a high-resolution, high-contrast 4,096 grayscale image. |
Seamless Network Integration
Ethernet 10/100/1000 Base T connectivity with DICOM 3.0 compatibility enables seamless data transfer to any DICOM hard copy output device, PACS, or HIS/RIS for efficient printing, archiving, and remote viewing of images. The system complies with IHE (Integrating the Healthcare Enterprise) standards for interoperability to ensure easy, effective integration with the hospital network. |
|
|
Back to Top |
Printers and Projectors
If your imaging requirements include color or black-and-white prints in a variety of sizes, Canon printers offer the perfect blend of print quality, easy-of-use, speed, and cost-effective operation. For teaching facilities and conference rooms, Canon’s high quality projectors provide LCOS (liquid crystal on silicon) technology, which is engineered to provide exceptional color, intricate details, and easy-to-read type.
For additional information, download the printer and projector brochures:
imagePROGRAF iPF500 Large Format Printer
PIXMA Pro9500 Printer
PIXMA Pro9000 Printer
REALiS SX7 Projector
For additional information, download the CXDI-50C
brochure.
For information on optional wireless network communications,
please Contact Us. |
|
Specifications |
Detector |
|
Detector |
LANMIT 7, Scintillator and Amorphous Silicon (a-Si) |
|
Scintillator |
Cesium Iodide |
|
Pixel Pitch |
160 microns |
|
Pixels |
2,208 x 2,688 (5.9 million) |
|
Imaging Area |
14 inches x 17 inches (35 cm x 43 cm) |
|
|
|
Image Acquisition and Processing |
|
Image Processing |
MLT(M) and MLT(S) are included |
|
A/D |
14-bit |
|
Grayscale |
12-bit (4,096 gradations) |
|
Preview Image |
3-5 seconds |
|
|
|
Data Output and Network Connection |
|
Network |
Ethernet 10/100/1000 Base T, RJ45 standard, wireless optional |
|
Data Output |
DICOM 3.0 compliant, Print Management Service Class (SCU), Storage Service Class (SCU), and others |
|
|
|
Electrical and Environmental |
|
Voltage |
100V, 120V, 230/240V (50/60Hz) |
|
Power Consumption |
200VA maximum (Detector unit only) |
|
Operating Environment |
41-95°F (5-36°C), 30-75% RH (non-condensing)
(Detector unit only) |
|
|
|
Physical Characteristics |
|
Dimensions WxLxH |
19.3 inches x 18.8 inches x 0.9 inches
(491 mm x 477 mm x 23 mm)
(Detector unit only) |
|
Weight |
10.6 pounds (4.8 kg) |
| |
|
Back to Top |
|