| CXDI-31
|
 |
|
| |
Introduction |
The CXDI-31 is the world’s first portable
DR system. With a compact and lightweight design,
the Canon CXDI-31 can be freely positioned to capture
images at angles difficult to get with fixed devices
such as lateral and axial imaging of shoulders, skull,
neck and extremities. The high-resolution CXDI-31
is perfect for pediatric and orthopedic applications. |
| |
Features |
As the world’s first portable digital radiography
detector, the CXDI-31 extends the benefits of DR to
a wide range of applications, including neonatal and
trauma radiography.
• Compact and Lightweight
The compact size of the CXDI-31 provides image capture
from virtually any angle or position. Yet the detector
weighs only 6.2 pounds!
• Enhanced Productivity
The technologist selects the necessary views to
complete the requested exam or the requested views
are automatically selected through communication
between the DICOM worklist and the RIS.
• Immediate Results
An on-screen preview image is available just 3-5
seconds after exposure. Sequential images can be
acquired in as little as 6 seconds.
• Seamless Network Integration
Ethernet 10/100 Base T connectivity with DICOM 3.0
compatibility enables seamless data transfer to
any DICOM hard copy output device, PACS, or HIS/RIS
for efficient printing, archiving, and remote viewing
of Images. The system complies with IHE (Integrating
the Healthcare Enterprise) standards for interoperability
to ensure easy, effective integration with the hospital
network.
For additional information, download the CXDI-31
brochure.
|
Specifications |
|
| Detector |
|
| |
Detector |
LANMIT 2, Amorphous Silicon and Scintillator |
| |
Scintillator |
Gadolinium Oxysulfide (GOS) |
| |
Pixel Pitch |
100 x 100 microns |
| |
Pixels |
2,256 x 2,878 (6.4 million) |
| |
Imaging Area |
9” x 11” (23 cm x 29 cm) |
| |
| Image Acquisition
and Processing |
|
| |
A/D |
14 bit |
| |
Grayscale |
12 bit (4,096 gradations) |
| |
Preview Image |
3-5 seconds |
| |
Refresh Cycle |
6-10 seconds |
| |
| Data Output and
Network Connection |
|
| |
Network |
Ethernet 10/100 Base T, RJ45 standard,
wireless optional |
| |
Data Output |
DICOM 3.0 compliant, Print Management Service Class
(SCU), Storage Service Class (SCU), and others |
| |
| Electrical and
Environmental |
|
| |
Voltage |
100V, 120V, 230/240V (50/60Hz) |
| |
Power Consumption |
300VA maximum (Detector unit only) |
| |
Operating Environment |
(Detector unit only)
50-95°F (10-35°C), 30-75% RH (non-condensing) |
| |
| Physical Characteristics |
|
| |
Dimensions WxLxH |
12.8” x 12.9” x 0.8” (324 x 327
x 20 mm) (Detector unit only) |
| |
Weight |
6.2 lbs. (2.8 kg) |
For additional information, download the DICOM