| CXDI-40EG
|
|
Introduction |
With multiple configurations designed for a wide
range of general radiography applications, Canon CXDI-40EG
can capture 17” X 17” X-ray images from
almost any position or angle. This flexible system
can be retrofit into a Bucky table, wall stand, or
universal stand. Versatile and economical, the Canon
CXDI-40EG’s compact, lightweight design is ideal
for upgrading existing radiography suites with digital
imaging capabilities.
|
| |
Features |
The high-resolution CXDI-40EG detector features
an expansive dynamic range with state-of-the-art LANMIT
6 (Large Area New-MIS Sensor and TFT) technology to
achieve high-quality diagnostic images.
-
Large Imaging Area
The large 17” x 17” imaging area flexibly
accommodates both large and small format anatomical
views without the need to adjust detector orientation.
-
Enhanced Productivity
The technologist selects the necessary views to
complete the requested exam or the requested views
are automatically selected through communication
between the DICOM worklist and the RIS.
-
Immediate Results
An on-screen preview image is available just 3-5
seconds after exposure. Sequential images can
be acquired in as little as 6 seconds.
-
Seamless Network Integration
Ethernet 10/100/1000 Base T connectivity with
DICOM 3.0 compatibility enables seamless data
transfer to any DICOM hard copy output device,
PACS, or HIS/RIS for efficient printing, archiving,
and remote viewing of Images. The system complies
with IHE (Integrating the Healthcare Enterprise)
standards for interoperability to ensure easy,
effective integration with the hospital network.
For additional information, download the CXDI-40EG
brochure.
|
Specifications |
|
| Detector |
|
| |
Detector |
LANMIT 6, Amorphous Silicon and Scintillator |
| |
Scintillator |
Gadolinium Oxysulfide (GOS) |
| |
Pixel Pitch |
160 x 160 microns |
| |
Pixels |
2,688 x 2,688 (7.2 million) |
| |
Imaging Area |
17” x 17” (43 cm x 43 cm) |
| |
| Image Acquisition
and Processing |
|
| |
A/D |
14 bit |
| |
Grayscale |
12 bit (4,096 gradations) |
| |
Preview Image |
3-5 seconds |
| |
Refresh Cycle |
6-10 seconds |
| |
| Data Output and
Network Connection |
|
| |
Network |
Ethernet 10/100/1000 Base T, RJ45
|
| |
Data Output |
DICOM 3.0 compliant, Print Management Service Class
(SCU), Storage Service Class (SCU), and others |
| |
| Electrical and
Environmental |
|
| |
Voltage |
100V, 120V, 230/240V (50/60Hz) |
| |
Power Consumption |
200VA maximum (Detector unit only) |
| |
Operating Environment |
(Detector unit only)
41-95°F (5-36°C), 30-75% RH (non-condensing)
|
| |
| Physical Characteristics |
|
| |
Dimensions WxLxH |
21.7” x 21.7”x2.7” (550 x 550
x 67.5 mm) (Detector unit only) |
| |
Weight |
Approximately 44 lbs. (20 kg) without grid |
For additional information, download the DICOM
Conformance Statement / IHE Integration Statement.
|
|
 |
|
 |
 |